Modules
7-17
Pressure Monitoring
Setting Pressure Display
1. Touch the
Display
button until the desired selection is displayed. Average is an
average pressure reading; Systolic is the high pressure reading; Diastolic is the
low pressure reading.
Calibrating the Pressure Transducer
The pressure monitoring system requires each transducer to be calibrated to
establish a zero reference. When a pressure transducer requires zeroing, the icon
will display “CAL mmHg”.
To calibrate the pressure transducer:
1. Open pressure transducer to atmospheric pressure.
2. Select the
Pressure
subtab within the Safety tab or touch the pressure icon.
3. Ensure the correct device name is displayed.
4. Touch the
Zero
button to calibrate pressure system. Calibration takes a few
seconds. The button turns dark gray during calibration.
•
The Pressure subtab and the pressure icon on the CCM will display “0 mmHg”
when the pressure system has been zeroed.
5. Close the pressure transducer to atmospheric pressure.
mmHg
CAL
Device Name
Zero Button
Safety
ART Pres
Average
Air
Next
Pressures
Zero
CAL mmHg
Display:
300
360
Alert:
Alarm:
Level
Limits (mmHg)
16:07:51
Safety
ART Pres
Average
Air
Next
Pressures
Zero
0 mmHg
Display:
300
360
Alert:
Alarm:
Level
Limits (mmHg)
16:07:51
mmHg
0
Summary of Contents for 801763
Page 2: ...Advanced Perfusion System 1 Operators Manual...
Page 3: ...ii...
Page 11: ......
Page 67: ......
Page 109: ......
Page 119: ......
Page 129: ......
Page 157: ......
Page 167: ......
Page 219: ...6 20...
Page 221: ......
Page 235: ......
Page 245: ......
Page 255: ......
Page 275: ......
Page 295: ......
Page 311: ......
Page 329: ...Additional Information 12 18 10 11 6 7 9 1 4 8 14 5 2 3 12 13 18 17 16 19 15 20...
Page 359: ...mmHg L min...