System Base
3-22
Snaps
Front
Tabs
Front
Latch
To Remove Cover:
1. Pull up the latch lever to open each latch.
2. Lift the cover off the base and out from under the roller pumps.
Cable Cover
(for the Terumo® Advanced Perfusion System 1, 801763 and 801764,
with serial numbers less than 1000)
To Install Cover:
1. Slide front part of cover under roller pumps on the base.
2. Place cover over rear cable management tray.
3. Apply pressure to the outer edges of the front of the tray until tabs snap into
place over the cable management tray.
4. Apply pressure to the rear of the tray to lock the cable cover into the snaps.
To Remove Cover:
1. Grasp the outer edges of the side of the cover firmly and lift.