![Rohde & Schwarz R&S UPP User Manual Download Page 872](http://html.mh-extra.com/html/rohde-and-schwarz/rands-upp/rands-upp_user-manual_1477869872.webp)
Cascading
R&S
®
UPP
856
User Manual 1411.1055.32 ─ 10
WAITING_FOR_SLAVES
Master is waiting for connection requests from slave devices.
MASTER_CASCADED
At least one slave has been successfully integrated in the cascade.
TRY TO CONNECT
Rapid flashing indicates that a connection is being set up, e.g. after power
‑
up or if the
cascade is incomplete because, for example, an instrument has not been switched on.
ERROR
Error condition. "ERROR" LED flashes.
If the ERROR LED remains lit, this indicates that an error has occurred during SCPI
control. If an error is no longer present following readout of the error queue using
SYST:ERR?, the red LED goes out.
8.4.1.3
Configuration of Slave Device
The programming routine using the "CASCADE/LAN RESET" button is visualized
using a constellation diagram. A flashing signal sequence at the "CASCADE LED"
informs the user of the respective instrument state. For a list of the states with the cor-
responding signal sequences, see
Chapter 8.4.1.4, "State Coding of Slave Device"
Configuration of a Cascade