![Rohde & Schwarz R&S UPP User Manual Download Page 249](http://html.mh-extra.com/html/rohde-and-schwarz/rands-upp/rands-upp_user-manual_1477869249.webp)
Instrument Functions
R&S
®
UPP
233
User Manual 1411.1055.32 ─ 10
The panels for the analog and digital generator are almost identical; the settings are
therefore described together. Any differences are indicated.
With an activated multichannel generator (requires optionR&S
UPP
‑
B8, multichannel
audio generator), individual signals can be configured for all selected channels using
the "Universal" generator function (see
on page 268). All
other generator functions cause perfectly identical signals on all selected channels.
5.6.1
Sine
The generator generates a sine-wave signal. Frequency and level are the same for
both channels.
Function
Selects the sine signal.
Remote command:
SINE
Dither
In the digital generator, a noise component with adjustable amplitude can be superim-
posed on the signal. The checkbox enables or disables this function.
The parameter fields for Dither do not appear in the analog generator.
Remote command:
Sweep Ctrl
Used to activate the sweep system. The frequency or level of the signal can be varied
within a selectable range from the start value to the stop value. If frequency or level is
selected as the x-axis, the measured values can be displayed as a curve in the Sweep
Graph graphical window.
Generator Signals
Summary of Contents for R&S UPP
Page 1: ...R S UPP Audio Analyzer User Manual User Manual Version 10 1411105532 ÅP ...
Page 34: ...Contents R S UPP 18 User Manual 1411 1055 32 10 ...
Page 630: ...Instrument Functions R S UPP 614 User Manual 1411 1055 32 10 Format ...
Page 631: ...Instrument Functions R S UPP 615 User Manual 1411 1055 32 10 Format ...