Instrument Functions
R&S
®
UPP
395
User Manual 1411.1055.32 ─ 10
Reference value is the
un
weighted input signal, band-limited to Frq Lim Upp (such
as R&S
UPL / UPD
without
weighting filter).
●
Scenario 2 – Prefilter
off
, function filter
on
:
Reference value is the
un
weighted
broadband
input signal.
●
Scenario 3 – Prefilter as weighting filter, function filter
off
:
Reference value is the weighted input signal, band-limited to Frq Lim Upp.
●
Scenario 4 – Prefilter as
weighting
filter, function filter
on
:
Reference value is the
weighted
,
broadband
input signal.
●
Scenario 5 – Prefilter as
lowpass
filter with cutoff frequency like Frq Lim Upp, func-
tion filter as
weighting filter
:
Reference value is the unweighted input signal, band-limited to "Frq Lim Upp " (as
with R&S
UPL / UPD with weighting filter).
For a description of selectable weighting filters, see
Chapter 5.41, "Weighting Filters"
Note:
The previous generation of instruments R&S UPL and R&S UPD did not allow a
prefilter to be selected. Consequently, only measurement scenarios 1 and 5 were pos-
sible with these instruments.
Note:
The upper band limit can be limited in the analog analyzer using the "Bandwidth"
parameter field, which has the same effect as an additional lowpass filter.
Remote command:
For a description of the filters and their setting parameters, see
Frq Lim Low
Used to enter the lower band limit for calculating the distortion voltage; this limit does
not affect calculation of the total RMS.
Remote command:
SENSe[1]:FREQuency:LIMit:LOWer
Frq Lim Upp
Used to enter the upper band limit for calculating the distortion voltage
and
the total
RMS value.
Note:
If the fundamental of the measurement signal is above "Frq Lim Upp", the refer-
enced measurements (THD+N, SINAD and Noise) are not to be performed because
the energy of the fundamental is filtered out in the reference value. An error message
appears in the user info line and the measured value is identified as invalid (---).
Remote command:
SENSe[1]:FREQuency:LIMit:UPPer
Fnct Settling
See section
Settling for Measurement Functions
for description of the settings and
associated remote control commands.
Distortion Measurements
Summary of Contents for R&S UPP
Page 1: ...R S UPP Audio Analyzer User Manual User Manual Version 10 1411105532 ÅP ...
Page 34: ...Contents R S UPP 18 User Manual 1411 1055 32 10 ...
Page 630: ...Instrument Functions R S UPP 614 User Manual 1411 1055 32 10 Format ...
Page 631: ...Instrument Functions R S UPP 615 User Manual 1411 1055 32 10 Format ...