Getting Started
R&S
®
UPP
80
User Manual 1411.1055.32 ─ 10
The available selection depends on the settings made in the combo boxes (6) and
(7). The entries are self
‑
explanatory.
Example: If the "Analog, balanced" output has been selected in the combo box (6)
for signal feed to the DUT, the measurement task "D/A Linearity" is
not
available as
this is intended only for a digital output.
Selection of a measurement task sets up the R&S UPP audio analyzer accordingly
without further action being required. With some measurement tasks, the measure-
ment must be started by clicking the "Start continuous measurement" or "Start sin-
gle measurement" button (13).
If a "Measurement Task" is selected in the quickstart panel, screens 1 to 4 display
the settings as well as the numeric and/or graphical displays for the selected mea-
surement tasks.
10. Entry fields for customizing the generator settings.
Example: Adjustment of the start and stop frequency for a frequency response
measurement.
11. Entry fields for customizing the analyzer settings.
Example: Switching over the input coupling from AC to DC.
12. Entry fields for customizing the graphical display.
Example: Changing the unit on the Y axis.
13.
Meaning of the buttons from left to right:
●
Start continuous measurement
●
Start single measurement
●
Pause / continue measurement
●
Switch to expert mode
: Access to the full range of functions of the audio ana-
lyzer.
●
Autoscale graph
: One
‑
off automatic scaling of the graphical display based on
the current minimum and maximum values of the curves or bars.
●
Set to reference
: The Y value of the curve is used as a reference value once
at the value 1000 Hz or 997 Hz (1000 Hz with analog measurements, 997 Hz
with digital measurements), provided that the Y axis of the graphical display is
shown in dBr. The curve is adjusted to 0 dBr at this point.
14.
When the audio analyzer is switched on, the quickstart panel is available in
screen 5 without it having to be called specifically.
3.3.4.4
Tool tips
All the controls in the quickstart panel have so
‑
called tool tips. If you hold the mouse
pointer over a control for approx. 1 second, a brief description of the function appears.
Brief Introduction to Operation
Summary of Contents for R&S UPP
Page 1: ...R S UPP Audio Analyzer User Manual User Manual Version 10 1411105532 ÅP ...
Page 34: ...Contents R S UPP 18 User Manual 1411 1055 32 10 ...
Page 630: ...Instrument Functions R S UPP 614 User Manual 1411 1055 32 10 Format ...
Page 631: ...Instrument Functions R S UPP 615 User Manual 1411 1055 32 10 Format ...