Instrument Functions
R&S
®
UPP
373
User Manual 1411.1055.32 ─ 10
Remote command:
Factor
Used to enter the factor by which the center frequency of the bandpass filter or band-
stop filter is to be higher than the generator frequency. Only visible in "Freq Mode Gen
Track".
Setting an integer factor greater than 1 will therefore allow an individual harmonic to be
measured. If the fundamental is to be selected, the factor must be set to 1.
Remote command:
5.21.4
Peak Measurement
Measurement of the peak value of any given signal, and the rise and fall of the mea-
sured value all occur without a delay. The measured value contains both AC and DC
components.
If a square wave is applied, it is limited with respect to bandwidth by the upper mea-
surement limit of the selected analyzer, and overshoots occur at the edges (Gibbs phe-
nomenon). The overshoots are also measured during the peak analysis, which means
that the returned measured value may be greater than the input peak. FS values
greater than 1 may be measured, especially in the digital range.
Chapter 5.21, "Level Measurements"
Function (Peak)
Selection of peak measurement.
Remote command:
PEAK
Level Measurements
Summary of Contents for R&S UPP
Page 1: ...R S UPP Audio Analyzer User Manual User Manual Version 10 1411105532 ÅP ...
Page 34: ...Contents R S UPP 18 User Manual 1411 1055 32 10 ...
Page 630: ...Instrument Functions R S UPP 614 User Manual 1411 1055 32 10 Format ...
Page 631: ...Instrument Functions R S UPP 615 User Manual 1411 1055 32 10 Format ...