Getting Started
R&S
®
UPP
104
User Manual 1411.1055.32 ─ 10
a) The basic settings for configuring the analyzer are again made in the "Analyzer
Config" panel.
These settings are made in the same way as shown in the previous example.
b) Next, the "Analyzer
Function" panel is activated.
The measurement function and measurement method are set in this panel.
First, the measurement function is selected by opening the parameter list and
selecting the item THD+N
&
SINAD.
c) In the next field "Meas Mode", various modes for distortion factor measurement
can be selected; the presetting "THD+N" is correct for this example.
d) With the THD+N measurement, the fundamental of the signal to be measured
is suppressed using a notch filter. The audio analyzer can automatically deter-
mine the fundamental from the spectrum and adjust the notch filter to it ("Fun-
damental Auto").
Filter and Equalizer are only required in exceptional cases.
e) The calculation for harmonics and noise components can, if necessary, be
restricted to within band limits which can be entered in the two fields "Freq Lim
Low" and "Freq Lim Upp".
All other parameter fields are not relevant for this example. This completes the
settings in the "Analyzer Config panel".
4. Step: Displaying the measurement results – Numeric measurement display
Introduction to Instrument Operation Using Examples
Summary of Contents for R&S UPP
Page 1: ...R S UPP Audio Analyzer User Manual User Manual Version 10 1411105532 ÅP ...
Page 34: ...Contents R S UPP 18 User Manual 1411 1055 32 10 ...
Page 630: ...Instrument Functions R S UPP 614 User Manual 1411 1055 32 10 Format ...
Page 631: ...Instrument Functions R S UPP 615 User Manual 1411 1055 32 10 Format ...