![Rohde & Schwarz R&S UPP User Manual Download Page 636](http://html.mh-extra.com/html/rohde-and-schwarz/rands-upp/rands-upp_user-manual_1477869636.webp)
Instrument Functions
R&S
®
UPP
620
User Manual 1411.1055.32 ─ 10
Black: Explanation of fields. These explanations are not stored in the sweep list.
Green: Fixed entries in ASCII format.
Red: Entries which change depending on the settings in the corresponding DispConfig
panel.
Table 5-12: Explanation of terms:
Identifier
Identifies the file as a limit file. During loading this identification is used to
decide whether the file concerned can be loaded.
Number of following X/Y
pairs
Self-explanatory
Offset ON/OFF
A comment which states whether, during the generation of a limit file from a
measured value sequence, an offset value was added (ON) or not (OFF) to the
Y data.
Offset value
A comment which states which offset value was added to the Y data during the
generation of a limit file from a measured value sequence.
X and Y unit
Depending on the selected x
‑
axis and the measurement function, a wide variety
of units can be entered here for the X / Y values. Examples for the X values:
Hz, °, s, V, FS or UI. Examples for the Y values: V, dBV, dBr, … V/Vr, °, s, db, %
etc.
X/Y pairs
X/Y pairs in the specified units.
Example of a limit data set:
UPV LIMIT FILE
9 # Datacount
# Offset ON
# Offset value = 10.0 dBV
Hz dBV
100 -70
711 -70
789 -45
875 -10
970 -10
1076 -10
1193 -30
1323 -70
2000 -70
Format
Summary of Contents for R&S UPP
Page 1: ...R S UPP Audio Analyzer User Manual User Manual Version 10 1411105532 ÅP ...
Page 34: ...Contents R S UPP 18 User Manual 1411 1055 32 10 ...
Page 630: ...Instrument Functions R S UPP 614 User Manual 1411 1055 32 10 Format ...
Page 631: ...Instrument Functions R S UPP 615 User Manual 1411 1055 32 10 Format ...