Instrument Functions
R&S
®
UPP
422
User Manual 1411.1055.32 ─ 10
"Value "
The RMS measurement is performed over a preset time which must
be entered in the parameter field below.
Numeric input of the measurement time. All desired signals (including
the reference signals) must have occurred within this time. Channels
with a longer delay time are labeled as invalid after this time expires
and the user info line displays a short note. All measurement results
are updated after this time. Thus for continuous measurement this
numerical value also determines the repetition rate of the measure-
ment.
Remote command:
SENSe[1]:FUNCtion:APERture:MODE
Fnct Settling
See
Chapter 5.38, "Settling Methods"
Auto Trigger Threshold Low
Negative percentage trigger threshold, relative to the nominal RMS value of the range
selected for this channel. For a threshold of -10%, therefore, the analog signal in the
range "200 mV RMS" would trigger at a peak value less than -28.28 mV.
The delay time of a channel is triggered, that is, determined up to the time when the
audio signal first exceeds the "high" threshold or first undershoots the "low" threshold.
Remote command:
SENSe[1]:LIPSync:AUDio:THReshold:LOW
Auto Trigger Threshold High
Positive percentage trigger threshold, relative to the nominal RMS value of the range
selected for this channel. For a threshold of +10%, therefore, the analog signal in the
range "800 mV RMS" would trigger at a peak value greater than 226.27 mV.
The delay time of a channel is triggered, that is, determined up to the time when the
audio signal first exceeds the "high" threshold or first undershoots the "low" threshold.
Remote command:
SENSe[1]:LIPSync:AUDio:THReshold:HIGH
Delay Measurement
Summary of Contents for R&S UPP
Page 1: ...R S UPP Audio Analyzer User Manual User Manual Version 10 1411105532 ÅP ...
Page 34: ...Contents R S UPP 18 User Manual 1411 1055 32 10 ...
Page 630: ...Instrument Functions R S UPP 614 User Manual 1411 1055 32 10 Format ...
Page 631: ...Instrument Functions R S UPP 615 User Manual 1411 1055 32 10 Format ...