Instrument Functions
R&S
®
UPP
595
User Manual 1411.1055.32 ─ 10
"activated"
The upper limit check is activated.
The parameter field below is used to enter an upper limit in the unit
selected in the Unit
parameter field.
An upper limit can be defined for the measurement result; every mea-
sured value is then compared with this limit.
If a relative unit is selected, the associated reference value (a few
parameter fields higher up) is used.
The upper limit is displayed as a value (bottom right) and as a red
limit marker in the bargraph.
As soon as a limit is overranged, the measurement result and bar-
graph change from green to red; if the limit violation no longer exists,
the measurement result and bargraph are displayed in green again.
A red measured value in the numeric display also indicates a limit vio-
lation:
In order to monitor limit violations over a long period of time without
having to watch the measurement result display continuously, the
upper limit is displayed permanently in red as soon as it is over-
ranged after the [START] key has been pressed.
[START] key or re
‑
entering the limit resets the hold function.
This hold function is not visible in the numeric display.
"deactivated"
The upper limit check is deactivated.
Remote command:
SENSe<n1>: <n1> = 1| 2 | 6 (function measurements | Input Monitor | Level Monitor) ;
LIMLower<n2>: <n2>= 1 | 2 (2 for two-channel instruments only)
Combi Display
Summary of Contents for R&S UPP
Page 1: ...R S UPP Audio Analyzer User Manual User Manual Version 10 1411105532 ÅP ...
Page 34: ...Contents R S UPP 18 User Manual 1411 1055 32 10 ...
Page 630: ...Instrument Functions R S UPP 614 User Manual 1411 1055 32 10 Format ...
Page 631: ...Instrument Functions R S UPP 615 User Manual 1411 1055 32 10 Format ...