Maintenance and Instrument Interfaces
R&S
®
UPP
880
User Manual 1411.1055.32 ─ 10
Figure 9-1: Pin assignment of IEC/IEEE-bus interface
9.4.1.2
IEC/IEEE-bus lines
●
Data bus with 8 lines DIO 1...DIO 8
Data is transferred parallel by bit and serial by byte in ASCII/ISO code. DIO1 is the
bit with the lowest value and DIO8 the bit with the highest value.
●
Control bus with 5 lines:
–
IFC
(Interface Clear):
Active LOW resets the interfaces of the connected instruments to their initial
setting.
–
ATN
(Attention):
Active LOW signals the transmission of interface messages
non-active HIGH indicates the transmission of device messages
–
SRQ
(Service Request):
Active LOW allows the connected instrument to send a service request to the
controller.
–
REN
(Remote Enable):
Active LOW allows switchover to remote-control mode.
–
EOI
(End or Identify):
This has two functions in conjunction with ATN:
ATN = HIGH Active LOW indicates the end of data transfer.
ATN = LOW Active LOW triggers a parallel poll.
●
3. Handshake bus with three lines:
–
DAV
(Data Valid):
Active LOW signals a valid data byte on the data bus.
–
NRFD
(Not Ready For Data):
Instrument Interfaces
Summary of Contents for R&S UPP
Page 1: ...R S UPP Audio Analyzer User Manual User Manual Version 10 1411105532 ÅP ...
Page 34: ...Contents R S UPP 18 User Manual 1411 1055 32 10 ...
Page 630: ...Instrument Functions R S UPP 614 User Manual 1411 1055 32 10 Format ...
Page 631: ...Instrument Functions R S UPP 615 User Manual 1411 1055 32 10 Format ...