Maintenance and Instrument Interfaces
R&S
®
UPP
885
User Manual 1411.1055.32 ─ 10
9.4.2.1
VXI-11 interface messages
With Ethernet connections, the interface messages are referred to as low-level control
messages. They can be used to emulate the interface messages of the IEC/IEEE bus.
Table 9-4: VXI-11 interface messages
Command
Effect on instrument
&ABO
(Abort)
Terminates processing of commands that have just been
received.
&DCL
(Device Clear)
Terminates processing of the commands that have just been
received, and sets the command-processing software to a
defined initial state. Does not modify the instrument configura-
tion.
>L
(Go to Local)
Transition to the "Local" state (manual operation).
>R
(Go to Remote)
Transition to the "Remote" state (remote control).
&GET
(Group Execute Trigger)
Triggers a measurement or a sweep.
&LLO
(Local Lockout)
The manual LOCAL switchover function is disabled.
&POL
(Serial Poll)
A serial poll is executed.
&NREN
(Not Remote Enable)
Manual switching to LOCAL mode is enabled.
9.4.3
USB Device Port
As standard, the instrument has a Type B USB device connector on the rear panel.
The RS
UPP can also be remote controlled via this port if the controller has the neces-
sary drivers.
Instrument Interfaces
Summary of Contents for R&S UPP
Page 1: ...R S UPP Audio Analyzer User Manual User Manual Version 10 1411105532 ÅP ...
Page 34: ...Contents R S UPP 18 User Manual 1411 1055 32 10 ...
Page 630: ...Instrument Functions R S UPP 614 User Manual 1411 1055 32 10 Format ...
Page 631: ...Instrument Functions R S UPP 615 User Manual 1411 1055 32 10 Format ...