![Rohde & Schwarz R&S UPP User Manual Download Page 450](http://html.mh-extra.com/html/rohde-and-schwarz/rands-upp/rands-upp_user-manual_1477869450.webp)
Instrument Functions
R&S
®
UPP
434
User Manual 1411.1055.32 ─ 10
Note:
If Undersample FFT is enabled, the waveform trigger source cannot be selected.
Instead, it is automatically set to Manual.
Selects the source (usually one of the measurement channels) which is to trigger mea-
surement recording if the trigger threshold is overranged or underranged.
Note:
In the case of multichannel analyzers, the trigger channel must be defined
before it can be used as trigger source for the waveform. If no trigger channel is activa-
ted, the trigger source is automatically set to "Manual".
"Channel 1"
Measurement channel 1 triggers recording.
Only available in the two
‑
channel analyzers if measurement channel
1 (or both channels) is activated.
"Channel 2"
Measurement channel 2 triggers recording.
Only available in the two
‑
channel analyzers if measurement channel
2 (or both channels) is activated.
"Trigger Chan"
The measurement channel selected as "Trigger Chan" triggers the
recording.
Only available in the multichannel analyzers if the trigger channel is
not deactivated.
Note:
If the trigger channel is deactivated, the "Trigger Source" is not
operable.
"Manual"
Recording is triggered as soon as the measurement starts.
"Gen Burst"
Recording is synchronized to the start of the generator signal (pref-
erably Sine Burst).
The generator signal is first muted briefly (approx. 3
ms) and then
restarted with zero
‑
degree phase ("Burst On" phase). Measurement
data recording is started at the same time, taking internal signal
delays into account.
It is thus possible to graphically determine the signal delay of the
DUT from the waveform display by selecting "Gen Burst" and using a
suitable signal from the internal generator (see
Note:
In this mode, the settling time of any filter that may be enabled
is not taken into account for starting the measurement. In other
words, the measurement starts without any delay and thus includes
the settling phase of the filters.
Analysis in the Time Domain
Summary of Contents for R&S UPP
Page 1: ...R S UPP Audio Analyzer User Manual User Manual Version 10 1411105532 ÅP ...
Page 34: ...Contents R S UPP 18 User Manual 1411 1055 32 10 ...
Page 630: ...Instrument Functions R S UPP 614 User Manual 1411 1055 32 10 Format ...
Page 631: ...Instrument Functions R S UPP 615 User Manual 1411 1055 32 10 Format ...