![Rohde & Schwarz R&S UPP User Manual Download Page 200](http://html.mh-extra.com/html/rohde-and-schwarz/rands-upp/rands-upp_user-manual_1477869200.webp)
Manual Operation
R&S
®
UPP
184
User Manual 1411.1055.32 ─ 10
4.17.1.2
Units for digital level measurement results
Basic unit: FS (Full Scale)
*) Level measurement result in hex notation
All digital analyzer measurement functions usually deliver values ranging from 0 to 1
FS.
FS values >1 may occur in the following cases:
●
In the case of the "Peak" measurement function with "Meas
Mode" "Peak
to
Peak"
●
In the case of a filter with gain (e.g. CCIR wtd) is used.
4.17.1.3
Units for distortion measurement results
Basic unit: %
Units