![Rohde & Schwarz R&S UPP User Manual Download Page 478](http://html.mh-extra.com/html/rohde-and-schwarz/rands-upp/rands-upp_user-manual_1477869478.webp)
Instrument Functions
R&S
®
UPP
462
User Manual 1411.1055.32 ─ 10
Remote command:
SENSe[1]:FUNCtion:SETTling:MODE
Samples
Designates the number of measured values to be covered by the settling algorithm, i.e.
it defines how many measured values are to be used in the settling measurement or
for averaging.
Note:
In this exceptional case, the term "sample" is used to designate an internal mea-
surement result instead of a digital sample.
Remote command:
SENSe[1]:FUNCtion:SETTling:COUNt
Tolerance
The tolerance value designates the maximum permissible value by which the current
measured value may differ from the preceding measured values so that it is still classi-
fied as settled.
With exponential settling, the selected tolerance value applies only to the difference
between the current and the last preceding measured value; for each earlier measured
value, the tolerance value increases by a factor of 2.
Remote command:
SENSe[1]:FUNCtion:SETTling:TOLerance
Resolution
The resolution value designates the maximum permissible absolute value by which the
current measured value may differ from the preceding measured values in order to be
accepted as settled, although the tolerance condition may be violated. This means that
a measured value is regarded as not settled only if it fails to meet both the tolerance
and the resolution condition.
With exponential settling, the selected resolution value applies only to the difference
between the current and the last preceding measured value; for each earlier measured
value, the resolution increases by a factor of 2.
Remote command:
SENSe[1]:FUNCtion:SETTling:RESolution
Timeout
Defines the maximum time that a settling algorithm is allowed to detect a settled result
after the start of a measurement.
Settling Methods
Summary of Contents for R&S UPP
Page 1: ...R S UPP Audio Analyzer User Manual User Manual Version 10 1411105532 ÅP ...
Page 34: ...Contents R S UPP 18 User Manual 1411 1055 32 10 ...
Page 630: ...Instrument Functions R S UPP 614 User Manual 1411 1055 32 10 Format ...
Page 631: ...Instrument Functions R S UPP 615 User Manual 1411 1055 32 10 Format ...