![Rohde & Schwarz R&S UPP User Manual Download Page 45](http://html.mh-extra.com/html/rohde-and-schwarz/rands-upp/rands-upp_user-manual_1477869045.webp)
Startup
R&S
®
UPP
29
User Manual 1411.1055.32 ─ 10
2.1.8
Status LEDs
[ERROR]
●
LED is lit (not flashing):
The instrument detected an error during the measure-
ment
●
LED is flashing:
The instrument detected an error during the configuration of the
cascading
[REMOTE]
The instrument is controlled remotely
[BUSY]
The LED is on as long as the measurement is running
[READY]
The instrument is ready for operation and can be controlled manually or remotely
[POWER]
●
LED is yellow:
The instrument is in standby mode
●
LED is green:
The instrument is switched on
[CASCADE]
●
LED is lit (not flashing):
The instrument is in cascading mode
●
LED is flashing:
The instrument is in the configuration phase of a cascading
[LAN]
The instrument is integrated in a local network
[GEN OVLD]
●
LED is yellow:
The outputs of the audio analyzer have been deactivated using the
"Output OFF" function.
●
LED is red:
The generator has been switched off
at high impedance
because an
external feed at pin 2 or 3 has caused a voltage of
≥
12 V peak to be applied to
instrument ground.
The outputs can be switched on again, either
●
by pressing CTRL+F12 on the external keyboard,
●
switching the instrument off/on with the standby mode pushbutton or
●
transmitting the SCPI command "OUTPut ON".
Description of the Front View
Summary of Contents for R&S UPP
Page 1: ...R S UPP Audio Analyzer User Manual User Manual Version 10 1411105532 ÅP ...
Page 34: ...Contents R S UPP 18 User Manual 1411 1055 32 10 ...
Page 630: ...Instrument Functions R S UPP 614 User Manual 1411 1055 32 10 Format ...
Page 631: ...Instrument Functions R S UPP 615 User Manual 1411 1055 32 10 Format ...