Instrument Functions
R&S
®
UPP
453
User Manual 1411.1055.32 ─ 10
"Freq & Sam-
ple Freq"
Only available for two-channel or multichannel measuring mode in
the digital analyzer.
The sampling frequency is measured on the channel last activated.
On all other channels (i.e. channel 1 in two-channel analyzers), a fre-
quency measurement is performed.
Note:
In this measuring mode, the audio frequency of the last chan-
nel is not measured.
The sampling frequency is measured with a resolution of 22.9
mHz
and an update interval of approx. 700
ms.
Note:
If the sampling frequency is changed by more than 3 Hz, a pre-
cision measurement is performed that takes max. 1.5 s. If the "Sam-
ple Freq" measurement is activated, therefore, the total time required
by a single measurement may be extended to up to 1.5 s.
"Sample Freq"
Only available in the digital analyzer.
The measurement windows contain the sampling frequency.
In this measuring mode, the audio frequency is not measured.
The sampling frequency is measured with a resolution of 22.9
mHz
and an update interval of approx. 700
ms.
Note:
If the sampling frequency is changed by more than 3 Hz, a pre-
cision measurement is performed that takes max. 1.5 s. If the "Sam-
ple Freq" measurement is activated, therefore, the total time required
by a single measurement may be extended to up to 1.5 s.
Remote command:
OFF | FREQuency | FQPHase | FQSamplefrequency |
SFRequency
Meas Time
Used to select the measurement speed.
Frequency, Phase and Group Delay Measurement
Summary of Contents for R&S UPP
Page 1: ...R S UPP Audio Analyzer User Manual User Manual Version 10 1411105532 ÅP ...
Page 34: ...Contents R S UPP 18 User Manual 1411 1055 32 10 ...
Page 630: ...Instrument Functions R S UPP 614 User Manual 1411 1055 32 10 Format ...
Page 631: ...Instrument Functions R S UPP 615 User Manual 1411 1055 32 10 Format ...