Instrument Functions
R&S
®
UPP
527
User Manual 1411.1055.32 ─ 10
Negative reference values < -1.0E-12 are computed using their magnitude, that is, they
are made into positive values.
As a result, trace displays in the referenced unit dBr, for instance, at a reference value
of 0.0 V have very high values for the upper limit, since the trace data is computed
internally using the reference value 1.0E
‑
12.
Available in the graphics subsystems
Sweep, FFT, Waveform, Bar Graph
Remote command:
DISPlay:SWEep<n2>:A:REFerence:VALue
DISPlay:SWEep<n2>:B:REFerence:VALue
Reference File
File name for the reference parameter File.
Available in the graphics subsystems
Sweep, FFT, Waveform, Bar Graph
Remote command:
DISPlay:SWEep<n2>:A:REFerence:FILE
DISPlay:SWEep<n2>:B:REFerence:FILE
Normalize
If a referenced unit (e.g. dBr or V/Vr) and a floating reference value, as in Reference =
Meas Ch1, Meas Ch2, Gen Track, or a reference trace held in a file, as in "File" or "File
Intern", are used, the displayed trace is normalized with the aid of the above by shifting
the whole curve, that is, multiplying it by the normalization value.
The normalization value is transferred to all scan data sets; the same normalization
value therefore applies to all scans.
Curve, Spectrum and Bar Graph Display
Summary of Contents for R&S UPP
Page 1: ...R S UPP Audio Analyzer User Manual User Manual Version 10 1411105532 ÅP ...
Page 34: ...Contents R S UPP 18 User Manual 1411 1055 32 10 ...
Page 630: ...Instrument Functions R S UPP 614 User Manual 1411 1055 32 10 Format ...
Page 631: ...Instrument Functions R S UPP 615 User Manual 1411 1055 32 10 Format ...