Instrument Functions
R&S
®
UPP
208
User Manual 1411.1055.32 ─ 10
Remote command:
Audio Bits
If the word width is decreased, the values of the audio samples are rounded to the
specified word width. Unused data bits are encoded as zeroes in the digital audio sig-
nal.
Value range: 8 ... 24
Remote command:
Sample Frequency
Used to set the clock rate of the digital audio signal for the generator.
The sample frequency setting is used as a reference whenever audio data signals are
generated.
The selected clock rate determines the maximum signal frequency which can be input.
For sine-wave signals, it is 49.9
% of the sampling rate.
If the sample frequency entered does not match the actual clock rate (e.g. with external
synchronization), all audio data signals shift accordingly in frequency.
This setting is also used to select the correct range for the Sync PLL.
Value range for numeric input:
30
kHz ... 200
kHz
Standard sampling frequencies can be set directly in the selection field.
Synchronization ranges for the generator sync PLL:
Base rate range
30
kHz ... 64
kHz
High rate range
65
kHz ... 128
kHz
Extended rate range
129
kHz ... 200
kHz
"32
...192
kHz" Standard sampling rate 32
kHz ... 192
kHz
"Value "
Variable input of sampling rate
Remote command:
Configuring the Digital Generator
Summary of Contents for R&S UPP
Page 1: ...R S UPP Audio Analyzer User Manual User Manual Version 10 1411105532 ÅP ...
Page 34: ...Contents R S UPP 18 User Manual 1411 1055 32 10 ...
Page 630: ...Instrument Functions R S UPP 614 User Manual 1411 1055 32 10 Format ...
Page 631: ...Instrument Functions R S UPP 615 User Manual 1411 1055 32 10 Format ...