Basic Operation
R&S
®
Cable Rider ZPH
61
User Manual 1321.0950.02 ─ 03
Entering the frequency
To get the best results, also enter the frequency of the signal under test.
1. Press the "Freq" softkey.
The R&S Cable Rider ZPH opens an entry box to enter the frequency.
2. Enter the frequency of the signal.
3. Confirm the entry with one of the unit keys.
The R&S Cable Rider ZPH transfers the new frequency to the internal power meter
which then corrects the measured power readings.
Activating the signal source
The R&S Cable Rider ZPH comes with a built-in signal source which you can enable or
disable in the power meter mode.
See
Once the signal source is enabled, you can output a signal source of -10 dBm at the
designated frequency from the RF output connector.
► Select the "SigGen On" softkey.
The R&S Cable Rider ZPH output a signal source of -10 dBm at the RF output con-
nector.
Using Power Meter