Basic Operation
R&S
®
Cable Rider ZPH
65
User Manual 1321.0950.02 ─ 03
Danger of skin burns and damage to the instrument
●
Never exceed the permissible continuous power.
●
See diagram on the rear of the sensor for the permissible continuous power.
●
Turn off the RF power to connect the sensor.
●
Screw the RF connectors tightly.
Test setup
Connect the power sensor cable to the USB port of R&S Cable Rider ZPH. If the power
sensor is having the binder connector (i.e R&S FSH-Z14, R&S FSH-Z44), the FSH-
Z144 adaptor cable is needed. Insert the directional power sensor between the source
and the load.
The power sensors for the R&S Cable Rider ZPH have an asymmetrical design.
Hence, you have to insert them into the test setup in such a way that the "Forward"
arrow (1
➙
2) on the sensor points toward the load (= in the direction of the power flux).
1 = Directional power sensor R&S FSH-Z14 or Z44
2 = Source
3 = Load
4 = USB binder adaptor (R&S FSH-Z144)
5 = USB port connector
Measuring the power
1. Press the MODE key.
2. Select the "Power Meter" softkey.
As soon as the R&S Cable Rider ZPH recognizes the power sensor, it displays the
type of the directional power sensor that is connected in the
onds, it also shows the forward power and return loss currently measured at the load.
Zeroing the power sensor
Before performing the power measurement, you should zeroing the power sensor.
Using Power Meter