![Rohde & Schwarz Cable Rider ZPH User Manual Download Page 237](http://html1.mh-extra.com/html/rohde-and-schwarz/cable-rider-zph/cable-rider-zph_user-manual_1477802237.webp)
Remote Commands
R&S
®
Cable Rider ZPH
202
User Manual 1321.0950.02 ─ 03
CALCulate<1...2>:DTF:CABLe:PRESet
<CableModel>
This command selects the cable model.
The numeric suffix at CALCulate is irrelevant for this command.
Suffix:
<1...2>
.
1...2
Parameters:
<CableModel>
String containing the filename of the cable model.
Example:
CALC:DTF:CAB:PRES '5088-HLFR.CBLMOD'
Selects the cable model from the file 5088-HLFR.CBLMOD
Selecting the Measurement Mode
The following commands select the measurement mode.
List of commands
●
MEASurement<1...2>:MODE
<MeasMode>
This command selects the measurement mode.
The numeric suffix at MEASurement is irrelevant for this command.
Suffix:
<1...2>
.
1...2
Parameters:
<MeasMode>
DTFault | REFLection | LOSS
DTFault
Selects the distance-to-fault measurement.
REFLection
Selects the reflection measurement (S11).
LOSS
Selects the cable loss measurement.
*RST:
REFLection
Example:
MEAS:MODE LOSS
Activates cable loss measurement.
Selecting the Measurement Format
The following commands select the measurement format in CAT mode.
List of commands
●
CALCulate<1...2>:TRACe:VSWR:MAX[:STATe]
●
Remote Control - Commands