Remote Commands
R&S
®
Cable Rider ZPH
222
User Manual 1321.0950.02 ─ 03
Usage:
Event
MMEMory:NAME
<FileName>
This command defines the path and filename that the R&S Cable Rider ZPH uses for
storing screenshots (see
on page 221). The path and filename
comply with DOS conventions.
This command is an event and therefore has no *RST value and no query.
Parameters:
<FileName>
String containing the filename.
Example:
MMEM:NAME 'Public\Screeshots\Test.png'
Stores the screenshot in the corresponding directory on the R&S
Cable Rider ZPH.
11.6.6 Configuring Data Capture
The following commands configure the data capture.
These commands are independent from the operating mode.
List of commands
●
●
SYSTem:CAPTure:DATaset[:STATe]
●
●
SYSTem:CAPTure:COUNter
<CaptureCounter>
This command defines the start of the filename counter.
The counter numbers the files stored when you capture data (screenshots, datasets
etc.).
Parameters:
<CaptureCounter>
String containing the number with which to start numbering files.
*RST:
'0000'
Example:
SYST:CAPT:COUN '0100'
Starts numbering files with 0100, e.g. Measurement0100.png.
SYSTem:CAPTure:DATaset[:STATe]
<State>
This command includes or excludes datasets from the data capture.
Parameters:
<State>
ON | OFF
*RST:
OFF
Remote Control - Commands