Remote Commands
R&S
®
Cable Rider ZPH
196
User Manual 1321.0950.02 ─ 03
Example:
CALC:MARK:X:SLIM ON
CALC:MARK:X:SLIM:LEFT 10MHz
CALC:MARK:X:SLIM:RIGH 100MHz
Turns on search limits and defines a search range from 10 MHz
to 100 MHz.
CALCulate<1...2>:MARKer<1...6>:Y?
This command queries the absolute vertical position of a marker.
If necessary, the corresponding marker is activated first.
To get a valid result, you have to perform a complete sweep with synchronization to
the sweep end between activating the delta marker and reading out the result. This is
only possible in single sweep mode.
The unit of the return value depends on
UNIT<1...2>:POWer
.
The numeric suffix at CALCulate selects the trace. The numeric suffix at MARKer
selects the marker.
Parameters:
<MarkerPosition>
Numeric value of the marker position.
Example:
INIT:CONF OFF
CALC:MARK2 ON
Turns on single sweep mode and marker 2.
INIT;*WAI
Starts a sweep and waits for the end.
CALC:MARK2:Y?
Queries the position of marker 2.
Usage:
Query only
11.6.2.6
Using Limit Lines
The following commands define limit lines and perform the corresponding limit checks.
List of commands
●
CALCulate<1...2>:LIMit<1...2>:BEEP[:STATe]
●
CALCulate<1...2>:LIMit<1...2>:COMMent?
●
CALCulate<1...2>:LIMit<1...2>:DELete
●
CALCulate<1...2>:LIMit<1...2>:FAIL?
●
CALCulate<1...2>:LIMit<1...2>:STATe
●
CALCulate<1...2>:LIMit<1...2>:UNIT:X?
●
CALCulate<1...2>:LIMit<1...2>:UNIT[:Y]?
●
CALCulate<1...2>:LIMit<1...2>:LOWer:SELect
●
CALCulate<1...2>:LIMit<1...2>:UPPer:SELect
●
CALCulate<1...2>:LIMit<1...2>:LOWer:THReshold
Remote Control - Commands