Remote Commands
R&S
®
Cable Rider ZPH
178
User Manual 1321.0950.02 ─ 03
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
FREQ:SPAN:AUTO ON
Turns on the automatic span determination.
[SENSe:]FREQuency:STARt
<StartFrequency>
This command defines the start frequency for measurements in the frequency domain
(span > 0).
Parameters:
<StartFrequency>
Range:
Depends on the operating mode and is specified in
the datasheet.
*RST:
0 Hz
Default unit: GHz
Example:
FREQ:STAR 20MHz
Defines a start frequency of 20 MHz.
[SENSe:]FREQuency:STOP
<StopFrequency>
This command defines the stop frequency for measurements in the frequency domain
(span > 0).
Parameters:
<StopFrequency>
Range:
Depends on the operating mode and is specified in
the datasheet.
*RST:
fmax
Default unit: GHz
Example:
FREQ:STOP 2000MHz
Defines a stop frequency of 2 GHz.
CALCulate<1...2>:DTF:DISTance:STARt
<DtfDistanceStart>
This command defines the start distance of the cable measurement.
The numeric suffix at CALCulate is irrelevant for this command.
Suffix:
<1...2>
.
1...2
Parameters:
<DtfDistanceStart>
Numeric value in the range from 3 m (10 ft) to 1500 m (4921 ft)
The unit is either meter or feet, depending on your selection.
*RST:
3 m (10 ft)
Default unit: m
Example:
CALC:DTF:DIST:STAR 50m
Sets the starting point of the measurement to 50 m.
Remote Control - Commands