Remote Commands
R&S
®
Cable Rider ZPH
183
User Manual 1321.0950.02 ─ 03
DISPlay<1...2>:VSWR:Y:SCALe:MAXimum
<DisplayRange>
This command defines the top value of the vertical axis for the VSWR measurement
format.
The numeric suffix at DISPlay is irrelevant for this command.
Suffix:
<1...2>
.
1...2
Parameters:
<DisplayRange>
Numeric value in the range from 1.1 to 71.
*RST:
21
Example:
DISP:VSWR:Y:SCAL:MAX 25
Defines a top value of 25 for the vertical axis.
DISPlay<1...2>:VSWR:Y:SCALe
<DisplayRange>
This command defines the bottom value of the vertical axis for the VSWR measure-
ment format.
The numeric suffix at DISPlay is irrelevant for this command.
Suffix:
<1...2>
.
1...2
Parameters:
<DisplayRange>
Numeric value in the range from 1.1 to 71.
The number you enter is rounded up to the next possible display
range. For example, if you enter 5, the R&S Cable Rider ZPH
automatically sets the display range to 1...6.
*RST:
1...21
Example:
DISP:VSWR:Y:SCAL 50
Defines a display range of 1...71.
11.6.2.3
Performing and Triggering Measurements
The following commands control the actual measurement process, including trigger
functionality.
Performing the Measurement
The following commands configure the sweep.
List of commands
●
●
INITiate:CONTinuous
<State>
This command selects the sweep mode.
Remote Control - Commands