Remote Commands
R&S
®
Cable Rider ZPH
186
User Manual 1321.0950.02 ─ 03
Parameters:
<MemorySlot>
Example:
CALC:MATH:COPY:MEM
Usage:
Setting only
Copies the trace into the memory.
CALCulate<1...2>:MATH<1...2>:STATe
<State>
This command turns trace mathematics on and off.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
MATH selects the number of the trace.
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
CALC:MATH:STAT ON
Switches on the trace mathematics.
DISPlay<1...2>[:WINDow]:TRACe<1...2>:MEMory[:STATe]
<State>
This command turns the memory trace on and off.
The numeric suffix at DISPlay is irrelevant for this command. The numeric suffix at
TRACe selects the number of the trace.
Parameters:
<State>
ON | OFF
*RST:
OFF
Example:
DISP:TRAC:MEM ON
Activates the memory trace.
DISPlay<1...2>[:WINDow]:TRACe<1...2>:MODE
<DisplayMode>
This command selects the trace mode.
If you are using the average, max hold or min hold trace mode, you can set the number
of measurements with
[SENSe:]SWEep:COUNt<SweepCount>
. Note that synchroni-
zation to the end of the average count is possible only in single sweep mode.
The numeric suffix at DISPlay is irrelevant for this command. The numeric suffix at
TRACe selects the number of the trace.
Remote Control - Commands