Remote Commands
R&S
®
Cable Rider ZPH
187
User Manual 1321.0950.02 ─ 03
Parameters:
<DisplayMode>
WRITe | AVERage | MINHold | MAXHold | VIEW | FREeze |
INFinite
*RST:
WRITe
You can turn off the trace with
DISPlay<1...2>[:WINDow]:TRACe<1...2>[:STATe]
.
For more information, see chapter "Trace Mode".
Example:
SWE:CONT OFF
SWE:COUN 16
Turn on single sweep mode and sets the number of measure-
ments to 16.
DISP:TRAC:MODE MAXH
Activates MAXHold mode for the trace.
INIT;*WAI
Starts the measurement and waits for the end of the 16 sweeps.
FORMat[:DATA]
<Format>[, <>]
This command selects the data format that is used for transmission of trace data from
the R&S Cable Rider ZPH to the controlling computer.
Note that the command has no effect for data that you send to the R&S Cable Rider
ZPH. The R&S Cable Rider ZPH automatically recognizes the data it receives, regard-
less of the format.
Parameters:
<Format>
ASCii | REAL
<>
ASCii
Returns the data in ASCII format, separated by commas.
REAL
Returns the data as 32-bit IEEE 754 floating point numbers in
the "definite length block format".
*RST:
ASCii
In REAL, 32 format, a string of return values would look like:
#42424<value 1><value 2>...<value n>
with
#4 representing the number of digits of the following number of
data bytes (= 4 in this example);
2524 representing the number of following data bytes (2524,
corresponds to the 711 sweep points of the R&S Cable Rider
ZPH;
<value> representing 4-byte floating point value.
Example:
FORM ASC
Selects the ASCii data format.
TRACe<1...2>[:DATA]?
[<arg0>]
This command queries the trace data of the current measurement.
Remote Control - Commands