Remote Commands
R&S
®
Cable Rider ZPH
214
User Manual 1321.0950.02 ─ 03
Parameters:
<Return values>
The return values depend on the power sensor in use and the
selected suffix at FETCh.
Measurements with R&S FSH-Z1 or R&S FSH-Z18:
FETC1:PMET?
power in dBm.
FETC2:PMET?
n/a
Measurements with R&S FSH-Z14 or R&S FSH-Z44:
FETC1:PMET?
forward power in dBm.
FETC2:PMET?
reflected power in dBm.
Example:
FETC2:PMET?
Returns nothing for R&S FSH-Z1 / R&S FSH-Z18 and the reflec-
ted power for R&S FSH-Z14 / R&S FSH-Z44.
11.6.3.8
Selecting a Telecommunication Standard
These commands apply radio communication standards to measurements with the
power sensor.
Note that the selection of a standard is available only for the power sensors R&S FSH-
Z1, -Z14, -Z18 and Z-44.
List of commands
●
CALCulate<1...2>:PMETer:PRESet[:STATe]
●
CALCulate<1...2>:PMETer:PRESet:SELect
CALCulate<1...2>:PMETer:PRESet[:STATe]
<State>
This command turns on or off the use of a standard.
The numeric suffix at CALCulate is irrelevant for this command.
Parameters:
<State>
ON | OFF
Example:
CALC:PMET:PRES ON
Activates usage of a standard.
CALCulate<1...2>:PMETer:PRESet:SELect
<Standard>
This command selects the standard for power sensor measurements.
The numeric suffix at CALCulate is irrelevant for this command.
Parameters:
<Standard>
E.g. GSM | EDGE | WCDMA | CDMAOne | CDMA2000 | DVBT |
DAB | TETRA | USER
Example:
CALC:PMET:PRES:SEL GSM
Selects the GSM standard for power sensor measurements.
Remote Control - Commands