Remote Commands
R&S
®
Cable Rider ZPH
182
User Manual 1321.0950.02 ─ 03
Suffix:
<1...2>
.
1...2
Parameters:
<DisplayRange>
Numeric value in the range from 1 dB to 150 dB.
The number you enter is rounded up to the next possible display
range. For example, if you enter 9, the R&S Cable Rider ZPH
automatically sets the display range to 10.
*RST:
100 dB
Default unit: dB
Example:
DISP:MAGN:Y:SCAL 50 DB
Defines a display range of 50 dB.
DISPlay<1...2>:MAGNitude:Y:SPACing
<ScalingType>
This command selects the scaling of the vertical axis for the magnitude measurement
format.
The numeric suffix at DISPlay is irrelevant for this command.
Suffix:
<1...2>
.
1...2
Parameters:
<ScalingType>
LINear | LOGarithmic
LINear
Selects a linear scale (%).
LOGarithmic
Selects a logarithmic scale (dB).
*RST:
LOGarithmic
Example:
DISP:MAGN:Y:SPAC LIN
Selects linear scaling.
DISPlay<1...2>:VSWR:Y:SCALe:MINimum
<DisplayRange>
This command defines the bottom value of the vertical axis for the VSWR measure-
ment format.
The numeric suffix at DISPlay is irrelevant for this command.
Suffix:
<1...2>
.
1...2
Parameters:
<DisplayRange>
Numeric value in the range from 1.0 to 70.
*RST:
1.0
Example:
DISP:VSWR:Y:SCAL:MIN 3
Defines a bottom value of 3 for the vertical axis.
Remote Control - Commands