Cable and Antenna Test Mode
R&S
®
Cable Rider ZPH
111
User Manual 1321.0950.02 ─ 03
Detection of the calibration kit is by means of the USB interface connection to the R&S
Cable Rider ZPH. See
.
When the calibration kit is detected, the R&S Cable Rider ZPH displayed the informa-
tion in the
Figure 7-7: Calibration kit information on title bar
1. Press CAL key to activate the calibration procedure.
The R&S Cable Rider ZPH displayed the message "Calibration unit detected".
2. Connect the calibration kit, R&S ZN-Z103 to "RF OUT/ REFLECTION" port of the
R&S Cable Rider ZPH
3. Select "Continue" to start the calibration.
The R&S Cable Rider ZPH performs the full 1-port calibration for a "Short", "Open"
or a "Load" connection automatically.
After completed the calibration routine, the R&S Cable Rider ZPH displayed the mes-
sage "Calibration done!". The
now shows "(Cal)" to indicate a successful
calibration.
Figure 7-8: Calibration kit, R&S ZN-Z103
1 = RF input
2 = Calibration kit, R&S ZN-Z103
3 = USB interface of calibration kit, R&S ZN-Z103
Performing Cable and Antenna Measurements