Remote Commands
R&S
®
Cable Rider ZPH
206
User Manual 1321.0950.02 ─ 03
Parameters:
<Return values>
<NORMalized>
Full calibration
<APPRoximate>
Approximate calibration (caused by a change of attenuation, for
example): measurement uncertainty must be anticipated
Example:
CAL:STAT?
Queries the calibration status of the R&S Cable Rider ZPH.
Usage:
Query only
CALibration<1...2>:CONTinue?
This command resumes calibration after a calibration standard has been connected.
After the command has been sent, it returns instructions on what to do to continue the
calibration process and finish it successfully.
The numeric suffix at CALibration is irrelevant for this command.
This command is a query and therefore has no *RST value.
Suffix:
<1...2>
.
1...2
Example:
CAL:CONT?
Resumes calibration and would return, e.g.:
'Connect Load to Port 1'
Connect the load before resuming calibration.
Usage:
Query only
CALibration<1...2>:MODE?
This command queries the general current calibration state.
The numeric suffix at CALibration is irrelevant for this command.
This command is a query and therefore has no *RST value.
Suffix:
<1...2>
.
1...2
Parameters:
<Return values>
<0>
Not calibrated
<1>
Calibrated
Example:
CAL:MODE?
If current calibration state is not calibrated, a '0' is returned.
Usage:
Query only
Remote Control - Commands