Remote Commands
R&S
®
Cable Rider ZPH
198
User Manual 1321.0950.02 ─ 03
<X-scale>
ABSolute | RELative
Scale of the x-axis.
<Y-unit>
DB | DBM | DBMV | DBUV | V | VOLT | W | WATT | DBUVM |
DBUAM | V_M | W_M2 | VSWR
Unit of the y-axis.
<X0...X99>
Data points on the x-axis.
Note: a limit line may consist of up to 100 horizontal data points.
<Y0...Y99>
Data points on the y-axis.
Example:
CALC:LIM:DEF 'Line','Example',HZ,ABS,DBM,
10000000,-10, 10000000,0,20000000,0
Defines a limit line with three data points.
Usage:
Setting only
CALCulate<1...2>:LIMit<1...2>:DELete
This command deletes a limit line.
This command is an event and therefore has no *RST value and no query.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
LIMit selects the limit line.
Suffix:
<1...2>
.
1...2
<1...2>
1...2
Example:
CALC:LIM2:DEL
Deletes the second limit line.
CALCulate<1...2>:LIMit<1...2>:FAIL?
This command queries the result of a limit check.
To get a valid result, you have to perform a complete sweep with synchronization to
the sweep end before reading out the result. This is only possible in single sweep
mode.
The numeric suffix at CALCulate is irrelevant for this command. The numeric suffix at
LIMit selects the limit line.
Suffix:
<1...2>
.
1...2
<1...2>
1...2
Example:
INIT;*WAI
CALC:LIM1:FAIL?
Performs a measurement and queries the result of the check for
limit line 1.
Remote Control - Commands