Section 7.2 Detector Replacement Calibration
The procedures below use the following applications:
•
System Warm-Up
(C:\Program Files\GE Medical Systems\eXplore Utilities\WarmupApp.exe)
•
Calibration Tools
(C:\Program Files\GE Medical Systems\eXplore Utilities\Calibration.exe)
•
X-Ray and Motor Controller
. The
X-Ray and Motor Controller
(C:\Program Files\GE
Medical Systems\eXplore Utilities\X_Ray_Controller.exe)
•
transgen
program (C:\Program Files\GE Medial Systems\eXplore
Utilities\Calibration\transgen.exe)
7.2.1
Mechanically Adjust Detector
This mechanical adjustment ensures the detector and stage are correctly positioned. To perform a
detector adjustment:
1.
Access the
X-Ray and Motor Controller
program.
2.
Ensure the fields below are set to the following settings:
Enable Left Limit: checked off
Enable Right Limit: checked off
Camera:
Thomson 8M
Bin Mode:
1 x 1
Exposure Time:
3000 ms