32. Click the
Full Res b
utton. De-select the
Mini-volume
field. This allows you to name the resulting
reconstructed volume.
33. Press the
Recon
button. Name the file
1x1wire_cs_ss.vff
(or
2x2wire_cs_ss.vff
or
4x4wire_cs_ss.vff
depending on which verification scans are being performed)
34. View the cropped area in
MicroView
upon
completion.
35. Zoom in on individual wires. All 4 wires should be as round as possible.
If all of the wires are not round, the Centre of Rotation needs to be adjusted as follows:
a.
Exit
MicroView
.
b.
Click the
Advanced
button in the
Reconstruction Utility
window.
c.
Click the
Reconstruction
tab.
d.
Change the Centre of Rotation field. Make only small, incremental changes to this number
(e.g. change 1150 to 1150.5) then click
OK
.
e.
Click the
Recon
button in the
Reconstruction Utility
window.
f.
View the image in
MicroView
upon completion. If the image is improved, continue to
change the Centre of Rotation in the same direction (e.g. change 1150.5 to 1151). If the
image is worse, change the Centre of Rotation in the opposite direction (e.g. change
1150.5 to 1149.5).
g.
Continue to adjust the Centre of Rotation until the wires are as round as possible (although
all 4 wires are unlikely to be equally round).
Round Wire
Skewed Wire
Содержание Healthcare eXplore Locus SP
Страница 2: ...This page is intentionally left blank...
Страница 12: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 12...
Страница 13: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page13...
Страница 37: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page37...
Страница 228: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 228...