20. Record the line length (appears in the top left of the screen). Make a note of this value in the
Worksheet.
21. Repeat the above steps 9 to 20 for the D2 image, recording values in the Worksheet.
Make note of its line length in the Worksheet.
22. Access the
Calculators
tab in
Calibration Tools
and
enter the following in the Source-Detector Distance
Calculator:
width1:
enter line length of D1
width2:
enter line length of D2
mag. dist.: enter the distance the tool was moved back
(i.e. 100mm).
23. Press the
Update
button. The system provides the
Source-Detector distance.
Make note of this value (should be approximately 250) in the Worksheet.