GE H
EALTHCARE
D
IRECTION
5158821-100, R
EVISION
3
E
X
PLORE
L
OCUS
SP S
ERVICE
G
UIDE
Page 80
A) REMOVE ROTARY STAGE FIRST, THEN SPECIMEN SLIDE
1.
Disconnect cables (4) to specimen slide.
2.
Using 4mm hex key, remove four screws
securing gear box slide spacer to specimen
slide.
3.
Using the 4mm hex key ball driver and 8mm
socket, remove four screws and nuts securing
specimen slide to angle plate.
4.
Using 2.5mm hex key remove two screws and
lockwashers securing limit finger to specimen
slide.
5.
Using 3/32” hex key remove limit switch bracket
from specimen slide.
4mm
Angle plate
4mm
8mm
2.5mm
Содержание Healthcare eXplore Locus SP
Страница 2: ...This page is intentionally left blank...
Страница 12: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 12...
Страница 13: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page13...
Страница 37: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page37...
Страница 228: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 228...