BACK UP CRITICAL FILES TO DVDs
1.
Copy the following files to a DVD:
-
C:\translations1
-
all the transform and transform.dat files found in D: drive
-
the D:\calib folder and the D:\detector folder
-
the C:\Program Files\GE Medical Systems folder
-
the newly created
calibration_values.txt
file
-
all saved volumes (e.g. 1x1_wire_cs_360)
-
the C:\Registry Backup folder.
2.
Label the DVD with the name
Calibration Values
, and include the system serial number and
date.
3.
Create zip files containing the acquired images and support files created for each of the 3 wire
scans. Call the zip files
1x1.zip
,
2x2.zip
and
4x4.zip
. There is no need to backup the 'Volumes'
or 'Corrected' folder.
4.
Save the zip files on to a separate DVD and label the DVD with the name
Wire Scans
. Include
the system serial number and date.
5.
Make two copies of each DVD. File one set of DVDs in the system's Device History Record
folder located on the back of the system, and send the other to Service Engineering.
Service Engineering
Attn: Janice Chu
1510 Woodcock Street, Unit #7
London, ON N6H 5S1
Canada
Содержание Healthcare eXplore Locus SP
Страница 2: ...This page is intentionally left blank...
Страница 12: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 12...
Страница 13: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page13...
Страница 37: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page37...
Страница 228: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 228...