6.2.2
Right Comair Fan Removal / Replacement
INTRODUCTION
The unit is equipped with two fans, located on the
rear panel, and provide fresh/cooling air to unit. A
slightly different procedure is required for each fan.
TOOL LIST
Standard toolkit
SUPPLIES
None.
SAFETY CONSIDERATIONS
X-ray shielding contains lead and lead compounds,
chemicals known (to the State of California) to cause cancer and other reproductive harm. Wear
gloves when handling shielding and wash hands after handling.
PRECONDITIONS
•
Machine locked out. See 6.1.1 for details.
•
Remove right access panel. See 6.1.3 for details.
REMOVE
1.
Using Phillips screwdriver, remove four screws,
flatwashers, and finger guard securing fan to
panel.
2.
Using slotted screwdriver, disconnect ground wire,
and unplug power wires from fan.
Right
fan
Left
fan
Right fan is behind
shield