Page 89
2.
Window, that is to be discarded, is located
behind the collimator. Washer is between
window and collimator.
RE-INSTALL COLLIMATOR
1.
Ensure that collimator is re-installed so that
opening is ’square’, and not on an angle.
FOLLOW-ON PROCEDURES
None
Collimator
Window