GE H
EALTHCARE
D
IRECTION
5158821-100, R
EVISION
3
E
X
PLORE
L
OCUS
SP S
ERVICE
G
UIDE
Page 130
8.
Click
File
and choose
Write Registry Values
from the drop-down list (and enter the password
EvS8899
when prompted).
9.
Use
Windows Explorer
to find the following Transform files in D:\
Transform1
Transform2
Transform4
Transform1.dat
Transform2.dat
Transform4.dat
Copy these transform files to:
C:\Program Files\GE Medical Systems\Explore Utilities\Parameters\
10. Enable test protocols. Change 1x1wire.txt, 2x2wire.txt and 4x4wire.txt to
1x1wire.protocol
,
2x2wire.protocol
and
4x4wire.protocol
.
Files are located in C:\Program Files\GE Medical Systems\eXplore Utilities\Protocols\.