GE H
EALTHCARE
D
IRECTION
5158821-100, R
EVISION
3
E
X
PLORE
L
OCUS
SP S
ERVICE
G
UIDE
Page 231
-
Before cleanup take note of sharp note of sharp corners or objects that could cut the
gloves. If gloves tear, remove, wash hands thoroughly and re-glove.
-
Use cloth or paper towels along with cleaner, taking care not to splash the material.
-
Sanitize the area using common bleach diluted 10:1 or a product listed under Approved
Cleaning Products below. Clean any tools that came in contact with a body fluid.
-
Since viruses require moisture to remain active, dry the entire area using a heat gun or
hair dryer.
-
When confident the area is clean and dry, place used cleaning materials in a red biohazard
bag.
-
Remove gloves, turning them inside out, and put gloves in the biohazard plastic bag. Seal
and give the bag to appropriate personnel for proper disposal.
Approved Cleaning Products:
-
Common household bleach - dilute 10:1
-
Fullsan 128 Neutral Germicidal Detergent - dilute 1 oz to 1 gal Fuller Brush Company:
phone number 1-800-438-5537
-
Wavecide 6 Disinfectant Spray - use full strength Edwards Medical: phone number 1-800-
837-7000
Section 8.2
Preventive Maintenance Procedures
8.2.1
X-Ray Source
1.
Access the
X-Ray and Motor Controller
software and adjust tube voltage and current to:
kV:
70
µA:
20
2.
Click on the
Turn X-Ray On
button. A buzz will sound for approximately 8 seconds, X-Ray
indicator lights will be illuminated.
The
X-Ray and Motor Controller
software reports that the x-ray tube is turned on and
displays approximately 70 kVp and 20 µA under Power Status.
The CM LED will display 70 kVp ± 0.2kVP and 20
µA.
Содержание Healthcare eXplore Locus SP
Страница 2: ...This page is intentionally left blank...
Страница 12: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 12...
Страница 13: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page13...
Страница 37: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page37...
Страница 228: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 228...