X-Ray and Motor Controller
indicates Safety door is opened, at least one of the four interlock
switches is not closed properly.
Check the electronic wiring of the interlock using a multi-meter:
When all four interlocks are closed, pin 10 and pin 11 of connector 3 on the electronic flange
should be a short circuit.