Page 142
7.2.13
Follow Up
Upon completion, back up calibration values.
SAVE CALIBRATION SETTINGS
1.
Capture a screen shot of the calibration values found under the
Calib. Settings
tab in
Calibration Tools
.
2.
Call the file
calibration_values.jpg
3.
Save the document to both C: and D: drives.
4.
Copy D:\calib folder to C: drive.
5.
Copy C:\translations1 to D: drive.
BACK UP REGISTRY
6.
If needed, add the following folder to the C: drive:
Registry Backup
7.
Create a sub-folder under Registry Backup and label the folder with "today’s date" e.g.
yymmdd
8.
Click on the Microsoft
start
button in the bottom right corner of the screen, and choose
Run...