Page 209
10. Capture an image in
Fluoro
as follows:
-
click
Menu
at the top of the
Fluoro
screen and choose
Capture
from the drop-down menu
-
once the Capture pop-up is displayed, enter
1
in the Number of Frames to Capture field
-
enter
10
in the Number of Frames to Average
field
-
name the file
bright
in the Name of Series field
-
save the file to the directory: D:\calib
-
click the
Capture
button.
The progress of capture is shown in the Fluoro’s
title bar as
[saving...]
. Wait until the image is
captured.
11. Turn off the X-Ray Source and capture a dark
image in Fluoro. Call this image
dark
.
12. Mount the X,Y Centring ball tool on the stage
cone. Physically rotate the X,Y Centring Ball
tool so that 2 of the 4 adjustment screws are
exactly perpendicular to the detector (i.e. not
at an angle).
13. Turn the X-rays back on and wait a few minutes
for the Source to stabilize, then re-start
Fluoro
.
14. Press
m
in
Fluoro
to display the mouse position window.
15. Elevate the tool using the Elevation feature in
X-Ray and Motor
Controller
until the ball appears in the centre of the
Fluoro
display. The
stage should be close to the detector. Move the stage 70 mm toward the Source so that it is
positioned approximately halfway between the Source and detector.
16. Click the
Turn X-Ray Off
button.
17. Mount the Dial Indicator Mount and Precision Dial Indicator.
top view
Содержание Healthcare eXplore Locus SP
Страница 2: ...This page is intentionally left blank...
Страница 12: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 12...
Страница 13: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page13...
Страница 37: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page37...
Страница 228: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 228...