REPLACE
1.
Reverse of removal.
2.
Attach limit switch bracket to specimen slide.
3.
Attach limit finger to specimen slide.
4.
Attach specimen slide to angle plate.
5.
Attach gear box slide spacer to specimen.
6.
Use cable ties to secure cables.
FOLLOW-ON PROCEDURES
•
Replace rotary stage, perform calibration and secure. Stage Replacement calibration is found
in Chapter 7.
•
Re-calibrate limit switches.
•
Replace shielding, case and covers, and perform an "abridged" calibration (see Chapter 7).