23. Select
Short Scan
. This looks at the first 200 views instead of all 360 views. It will emphasize
any errors in calibration. Click OK.
24. Press the
Recon
button in the
Reconstruction Utility
window.
25. View the image in
MicroView
upon completion. The 4 wires should appear "square" to the
screen.
26. Select a region of interest that encompasses all 4 wires using the 7 and 8 keys in
MicroView
.
27. Select File>Save crop coordinates.
28. Exit
MicroView
.
29. Click the
Load Crop
button on the
Reconstruction Utility
window.
30. Press the
Recon
button. View the image in
MicroView
and verify the image is cropped
properly, then exit
MicroView
.
31. Change the Z-Size field on the Reconstruction Utility screen to
1
.
32. Click the
Full Res b
utton. De-select the
Mini-volume
field. This allows you to name the resulting
reconstructed volume.
33. Press the
Recon
button. Name the file
1x1wire_cs_ss.vff
(or
2x2wire_cs_ss.vff
or
4x4wire_cs_ss.vff
depending on which verification scans are being performed).
Содержание Healthcare eXplore Locus SP
Страница 2: ...This page is intentionally left blank...
Страница 12: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 12...
Страница 13: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page13...
Страница 37: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page37...
Страница 228: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 228...