Page 88
6.2.15
Collimator Window Removal
INTRODUCTION
The collimator window is a thin Lexan shield located at the front of the x-ray source, behind the
bronze-coloured metal collimator. The window can be access through the sliding door, with the case
on.
This window, if present, should be discarded. It gets yellow and brittle, and may affect image
quality.
TOOL LIST
Standard toolkit
SUPPLIES
None
SAFETY CONSIDERATIONS
The X-ray source opening contains beryllium. Do not touch it.
WARNING
DO NOT
TOUCH THE
OPENING OF
THE X-RAY
SOURCE WITH
COLLIMATOR
REMOVED.
PRECONDITIONS
•
Machine locked out. See 6.1.1 for details.
•
Filter wheel assembly removed. See 6.2.9 for details.
•
X-ray source front shielding removed. See 6.1.5 for details.
REMOVE
1.
Using 9/64” hex key, remove three screws from
collimator. Note position of collimator for later
re-installation.
Collimator