13. Rotate the ball 180° and confirm the Dial Indicator reads zero.
14. Rotate the ball 90°.
NOTICE
Equipment
Damage
Possible
15. Carefully adjust the 2 screws that are aligned with the Dial Indicator until the Dial Indicator
reads zero.
16. Rotate the ball 180° and verify that the Dial Indicator still reads zero.
Note:
Fine tuning may be required. Continue to make adjustments until the ball can rotate 360° without
the Dial Indicator reading moving more than .0002".
17. Once the ball has been centred, remove the Dial Indicator Mount and Precision Dial Indicator.
18. Select
Magnification
in the Motor field in the
X-Ray and Motor Controller
window and move
the ball 70 mm toward the detector. This should be 2mm from the limit switch.
19. Warm up the x-rays without using the
System Warm-up
application to avoid moving the stage.
To do this, use the
X-Ray and Motor Controller
program and do the following:
a.
Set the kV field to 80
b.
Set the µA field to 10
c.
Press the
Turn X-Rays
on button
d.
Wait 2 minutes
e.
Press the
Turn X-Rays
off button
f.
Set the kV and µA field to the next value in list below, and repeat steps c to e:
- wait 2 minutes @ 80kV, 45µA
- wait 10 minutes @ 80kV, 80µA
20. Capture the image in
Fluoro
and name it
capture
.
21. Correct and unwarp the captured image using
Calibration Tools
.
Содержание Healthcare eXplore Locus SP
Страница 2: ...This page is intentionally left blank...
Страница 12: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 12...
Страница 13: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page13...
Страница 37: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page37...
Страница 228: ...GE HEALTHCARE DIRECTION 5158821 100 REVISION 3 EXPLORE LOCUS SP SERVICE GUIDE Page 228...